MIL-PRF-27G
4.7.12.3 DC resistance and resistive unbalance. The dc resistance of the windings shall be measured at, or
corrected to 20°C. The resistive unbalance of center-tapped windings in percent [[(R1 - R2)/R1] x 100] shall be
computed. R1 and R2 are the resistances of each part of the winding, and R1 ≥ R2 . For resistances under 1 ohm,
measurements shall be made with a Kelvin bridge or equivalent.
4.7.12.4 Inductance and inductive unbalance. The inductance of the windings shall be measured at the specified
test voltage and frequency with the specified dc current applied (see 3.1). The inductive unbalance of center-tapped
windings in percent [[(L1 - L2)/L1] x 100] shall be computed. L1 and L2 are inductances of each part of the winding,
and L1 ≥ L2 .
4.7.12.5 Harmonic distortion. The transformer shall be terminated in its proper source and load impedance. A
sine-wave voltage of specified frequency shall be applied such that the specified output conditions are achieved and
the total harmonic distortion shall then be computed or measured (see 3.1).
4.7.12.6 Primary impedance (for qualification only). Transformer primary impedance shall be measured with all
normally loaded secondaries loaded with their specified impedances, and with specified dc currents flowing in the
windings. The resistance and reactance looking into the primary shall be measured at the specified input frequency
and voltage by a bridge or equivalent method approved by the Government (see 3.1).
4.7.12.7 Frequency response. Using the circuit shown on figure 3, with the specified values of impedance,
sufficient EMF from the source of the specified reference frequency should be applied to present the specified voltage
across the load impedance. Keeping the source EMF constant, the output voltage is then measured at the specified
frequencies (see 3.1). The rated dc currents shall flow through the windings. The measuring device shall be a
vacuum-tube voltmeter or equivalent instrument having high-impedance input circuit. The same type of instrument
shall be used to measure the source EMF and the output voltage. Frequency response, in decibels (dB), may be
determined for each instrument as follows:
Ef
dB = 20 log E
r
Where:
Er = reference output voltage.
Ef = output voltage at any other frequency.
A frequency spectrum instrument may be used as an alternate method of test.
4.7.12.8 Insertion loss. Transformers shall be tested in accordance with 4.7.12.8.1 or 4.7.12.8.2, as specified (see
3.1).
4.7.12.8.1 Insertion loss, for equal input and output impedance. Using the circuit shown on figure 4 with a
constant voltage, Eg, across the generator, the load voltage, with and without the transformer in the circuit, shall be
measured at the specified frequency (see 3.1). The insertion loss in dB, shall be calculated using the following
formula:
E1
I.L. = 20 log10 E
2
Where:
I.L. = Insertion loss (in dB) with (Eg = E1) constant
E1 = Voltage with the transformer not in the circuit at the specified frequency
E2 = Load voltage with the transformer in the circuit at the specified frequency
132p23" style="font-size:11px">23
Downloaded: 2014-09-24T13:27Z
For Parts Inquires submit RFQ to Parts Hangar, Inc.
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business